Classification and maximal permissible errors
 
 CLASSIFICATION OF WEIGHING INSTRUMENTS
Class: Verifying unit
[e]
Quantity of verifying units
[n=Max/e]
Minimal load
[Min]*
MinimalMaximal Lower threshold
Special 0,001 g e ** 50.000   100 e
High 0,001 g e 0,005 100 100.000 20 e
0,1 g e 5.000 100.000 50 e
Average 0,1 g e 2 g10010.000 20 e
5 g e 500 10.000 20 e
Normal
5 g e 100 1.000 10 e
   * minimal load is reduced to 5e for instruments In classes II and III and settling fees according to sending tariffs.
 ** because of uncertainty of standard masses the checking of balance with unit e < 1 mg is not performed normally.
 
 MAXIMAL PERMISSIBLE ERRORS FOR FIRST VERIFICATION OF WEIGHING INSTRUMENTS
Maximal Permissible Errors at First Verification : For loads [m] In verifying units [e]
Class Class Class Class
± 0,5 e 0 m 50.000 0 m 5.000 0 m 500 0 m 50
± 1 e 50.000 < m 200.000 5.000 < m 20.000 500 < m 2.000 50 < m 200
± 1,5 e 200.000 < m 20.000 < m 100.000 2.000 < m 10.000 200 < m 1.000
 Maximal errors for instruments in usage are two times higher from errors acceptable for first verification.
 
  