| Balance protective cover offered in standard |
|
MYA/2Y series of microbalances is designed to meet the highest requirements for determination of mass.
Measurement reliability and accuracy are maintained by system of automatic internal adjustment / calibration.
Microbalances comprise two major components (an electronic module and a precise mechanical measuring system are enclosed separately). Such design eliminates the influence of heat sourcing from instrument's electronics on its mechanical components and additionally protects it from shocks and vibrations caused by users operating the instrument.
All the elements of a microbalance are made of glass and steel which eliminates the influence of electrostatics on weighing process.
|
|
| |
|
 |
|
 |
|
 |
Additional weighing pan ø 60 mm
- designed for weighing filters,
- used interchangeably on a microbalance MYA 0,8/3. |
|
Infrared proximity sensors
Optional functions:
- PRINT function
- TARE function
- sensor's sensitivity adjustment |
|
Electronic level indicator
Functions:
- ALARM function
- graphic level indicator
- programmable acceptable tilts |
|
|
|
|
|
 |
|
 |
|
|
Data exchange through USB storage devices
- update balance software
- export weighing data
- export/import databases
- export/import balance settings
- exchang data between balances |
|
Communication interfaces
- Ethernet
- 2×RS 232
- 2×USB
- additional display port |
|
|
|
|
|
| Functions: |
 |
Filling |
 |
Checkweighing |
 |
Percent setup |
 |
Statistics |
 |
Formulation |
 |
Air
buoyancy
correction |
 |
GLP
procedures |
 |
Autotest |
 |
Infrared
sensors |
|
 |
3
years warranty period |
| |
| |
|
| Technical data: |
MYA 2/2Y |
MYA 0,8/3/2Y |
MYA 5/2Y |
MYA 11/2Y |
MYA 21/2Y |
 |
 |
 |
 |
 |
| Max capacity |
2 g |
0,8/3 g |
5 g |
11 g |
21 g |
| Readability |
1 µg |
1/10 µg |
1 µg |
1 µg |
1 µg |
| Tare range |
-2 g |
-3 g |
-5 g |
-11 g |
-21 g |
| Repeatability * |
1 µg |
1,6 µg |
2 µg |
| Linearity |
±3 µg |
±5 µg |
±6 µg |
±7 µg |
| Eccentric load deviation |
3 µg |
5 µg |
6 µg |
7 µg |
| Sensitivity offset |
1,5×10-6×Rt |
3×10-6×Rt |
4×10-6×Rt |
| Sensitivity temperature drift |
1×10-6/°C×Rt |
| Sensitivity stability |
1×10-6/Year×Rt |
| Minimum weight (USP) |
3 mg |
4,8 mg |
6 mg |
| Minimum weight (U=1%, k=2) |
0,2 mg |
0,32 mg |
0,38 mg |
| Pan size |
ø 16 mm |
ø 16 + ø 60 mm
(weighing pan for filters) |
ø 26 mm |
| Weighing chamber dimensions |
ø 90×90 mm |
| Stabilization time |
5 s |
| Adjustment/calibration |
internal (automatic) |
| Power supply |
110 ÷ 230 V AC / 50 ÷ 60 Hz / 13,5 ÷ 16 V DC / 1,1 A |
| Display |
5,7'' touch screen |
| Interface |
2×USB, 2×RS 232, Ethernet, 2Inputs/2Outputs |
| Net weight/Gross weight |
10,2/14,7 kg |
| Packaging size |
565×565×355 mm |
| Ambient conditions: |
| Working temperature |
+10° ÷ +40°C |
| Change rate of working temperatures |
±0,3 °C/h (±1 °C/8h) |
| Atmospheric humidity |
45% ÷ 65% |
| Change rate of atmospheric humidity |
±1%/h (±4%/8h) |
|