RADWAG BloggerRADWAG FacebookRADWAG RSS
 
MYA/2Y microbalances

Balance protective cover offered in standard

 

MYA/2Y series of microbalances is designed to meet the highest requirements for determination of mass. Measurement reliability and accuracy are maintained by system of automatic internal adjustment / calibration.

Microbalances comprise two major components (an electronic module and a precise mechanical measuring system are enclosed separately). Such design eliminates the influence of heat sourcing from instrument's electronics on its mechanical components and additionally protects it from shocks and vibrations caused by users operating the instrument.
All the elements of a microbalance are made of glass and steel which eliminates the influence of electrostatics on weighing process.


 
   
interfejs komunikacyjny
Additional weighing pan  ø 60 mm
- designed for weighing filters,
- used interchangeably on a microbalance MYA 0,8/3.
  Infrared proximity sensors
Optional functions:
- PRINT function
- TARE function
- sensor's sensitivity adjustment
  Electronic level indicator
Functions:
- ALARM function
- graphic level indicator
- programmable acceptable tilts
     
Data exchange through USB storage devices
- update balance software
- export weighing data
- export/import databases
- export/import balance settings
- exchang data between balances
  Communication interfaces
- Ethernet
- RS 232
- 2×USB
- additional display port
   
 
Functions:
Filling
Checkweighing
Percent setup
Statistics
Formulation
Air buoyancy
correction
GLP
procedures
Autotest Autotest
Infrared
sensors
    
3 years warranty period
 
Download:
Product folder
User manual
Declaration
of conformity
Air buoyancy
correction -
characteristics
 
Technical data: MYA 2/2Y MYA 0,8/3/2Y MYA 5/2Y MYA 11/2Y MYA 21/2Y
Max capacity 2 g 0,8/3 g 5 g 11 g 21 g
Readability 1 µg 1/10 µg 1 µg 1 µg 1 µg
Tare range -2 g -3 g -5 g -11 g -21 g
Repeatability * 1 µg 1,6 µg 2 µg
Linearity ±3 µg ±5 µg ±6 µg ±7 µg
Eccentric load deviation 3 µg 5 µg 6 µg 7 µg
Sensitivity offset 1,5×10-6×Rt 3×10-6×Rt 4×10-6×Rt
Sensitivity temperature drift 1×10-6/°C×Rt
Sensitivity stability 1×10-6/Year×Rt
Minimum weight (USP) 3 mg 4,8 mg 6 mg
Minimum weight (U=1%, k=2) 0,2 mg 0,32 mg 0,38 mg
Pan size ø 16 mm ø 16 + ø 60 mm
(weighing pan for filters)
ø 26 mm
Weighing chamber dimensions ø 90×90 mm
Stabilization time 5 s
Adjustment/calibration internal (automatic)
Power supply 110 ÷ 230 V AC / 50 ÷ 60 Hz / 13,5 ÷ 16 V DC / 1,1 A
Display 5,7'' touch screen
Interface 2×USB, 2×RS 232, Ethernet, 2Inputs/2Outputs
Net weight/Gross weight 10,2/14,7 kg
Packaging size 565×565×355 mm
Ambient conditions:
Working temperature +10° ÷ +40°C
Change rate of working temperatures ±0,3 °C/h (±1 °C/8h)
Atmospheric humidity 45% ÷ 65%
Change rate of atmospheric humidity ±1%/h (±4%/8h)
Rt - net weight
* repeatability is expressed as a standard deviation from 10 weighing cycles.
Accessories:
Antivibration table for microbalanceTHB 2 ambient conditions module
Professional weighing tableAdditional LCD display "WD-3/01/Y"
Kafka thermal printerPC keyboard
Dot matrix Epson printerAdditional adapter for pipettes calibration
Label printer CitizenPower adapter with battery and charger ZR-02
Anti draft shield for microbalancesMass standard
Air density determination kitAntistatic cable PA1
"Tare" or "Print" foot buttonBar code scanner
"PW-WIN" computer softwareCable RS 232 (scale - Kafka printer) "P0136"
"RAD-KEY" computer softwareCable RS 232 (scale - computer) "P0108"
"REC-FS" computer softwareCable RS 232 (scale, Epson, Citizen printer) "P0151"
Antistatic ionizer DJ-03 
Please click entry from additional equipment list to see details.
©2005 RADWAG Wagi Elektroniczne. All rights reserved.
All proprietor's rights do the materials placed in the server are in possession of RADWAG Wagi Elektroniczne and are
not to be copied, processed and utilized in commercial purpose unless agreed with RADWAG Wagi Elektroniczne.