RADWAG BloggerRADWAG FacebookRADWAG RSS
 
Precision balances APP/2C
 
Downloads:
Product folder
User manual
Declaration of conformity

Precision balances series APP/2C are offered with weighing platform dimensions 365×258mm. Balances feature backlit LCD display.

Independently on version, APP balance is based on electromagnetic measuring system, external adjustment/calibration system (2C version features internal adjustment/calibration system) and RS 232 output. Balance housing is made of aluminum, weighing pan is in stainless steel technology.

Balances series APP have under hook weighing as standard solution. It is an alternative solution for weighing loads with non-standard dimensions and those creating magnetic field. Under hook weighing is also applied for density determination procedure.

 

Function:
Parts counting
Density determination
Percent setup
 
      *
 
2 years warranty period

Weighing has just
become faster
 
Dane techniczne: APP 10/2C APP 25/2C APP 30/2C APP 35/2C -
- - - -
APP 10/2C/1 APP 25/2C/1 APP 30/2C/1 APP 35/2C/1 APP 50/2C/1
- - - - -
Max capacity 10 kg 25 kg 30 kg 35 kg 50 kg
Minimal load

0,5 g

5 g
Readability 0,01 g 0,1 g 0,1 g 0,1 g 0,1 g
Tare range -10 kg -25 kg -30 kg -35 kg -50 kg
Repeatability 0,01 g 0,1 g 0,1 g 0,1 g 0,15 g
Linearity ±0,02 g ±0,1 g ±0,3 g ±0,3 g ±0,5 g
Pan size 347×259 mm
Stabilization time 1,5 s   3 s    
Sensitivity drift 2ppm/°C in temperature +15 ° - +35 °C
Working temperature +10 ° - +40 °C
Power supply 110-230 V AC / 50 ÷ 60 Hz / 13,5 ÷ 16 V DC / 1,1 A
Adjustment/calibration internal (2C) or external (2C/1)
Display LCD (backlit)
Interface RS 232
Net weight/Gross weight APP/2C 19,9/22,9 kg
APP/2C/1 13,5/16,5 kg
Packaging size 570×560×325 mm
 
Dane techniczne: APP 6/35/2C
APP 6/35/2C/1
-
Max capacity 6/35 kg
Minimal load 5 g
Readability 1/5 g
Tare range -35 kg
Repeatability 1/5 g
Linearity ±1/5 g
Pan size 347×259 mm
Stabilization time 3 s
Sensitivity drift 2ppm/°C in temperature +15 ° - +35 °C
Working temperature +10 ° - +40 °C
Power supply 110-230 V AC / 50 ÷ 60 Hz / 13,5 ÷ 16 V DC / 1,1 A
Adjustment/calibration internal (2C) or external (2C/1)
Display LCD (backlit)
Interface RS 232
Net weight/Gross weight APP/2C 19,9/22,9 kg
APP/2C/1 13,5/16,5 kg
Packaging size 570×560×325 mm
Accessories:
Antivibration table for laboratory balancesCalibration weight
Kafka thermal printerMass standard
Dot matrix Epson printerPower loop module AP2-1 (plastic version)
"PW-WIN" computer softwareAntistatic cable PA1
"RAD-KEY" computer softwareStand for APP
"REC-FS" computer softwareCable RS 232 (scale - Kafka printer) "P0136"
Additional LCD display "WD-3/01"Cable RS 232 (scale - computer) "P0108"
Power adapter with battery and charger ZR-02Cable RS 232 (scale, Epson, Citizen printer) "P0151"
Please click entry from additional equipment list to see details.
©2005 RADWAG Wagi Elektroniczne. All rights reserved.
All proprietor's rights do the materials placed in the server are in possession of RADWAG Wagi Elektroniczne and are
not to be copied, processed and utilized in commercial purpose unless agreed with RADWAG Wagi Elektroniczne.